Open original FDA record ↗Recall record
- FDA identifier
- Z-0824-2011
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Dec 28, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm has determined that two lots of this instrument were inappropriately manufactured by the supplier. The manufacturing process created a step in the lumen which impedes the Lag Screw Retaining Shaft from passing through the lumen without the use of excessive force.
Source facts
- Action
- The firm sent Correction and Removal Notification Immediate Response Required letters via email to consignees. The letter identified the affected product and the reason for the recall. The letters also discussed the clinical implications and required actions. Customers are to complete the inventory certification form and user facility documentation form. Each of the affected accounts will be provided with a new instrument, manufactured with the correct lumen diameter and will secure and return the affected instruments per the attached notification, along with the completed inventory certification form. If any further information is needed, customers are to contact Aimee Wood at 574-372-4463.
- Initiated
- 2010-10-29
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-09-16
- Distribution
- Worldwide Distribution -- USA, including states of MI, NJ, MS, OR, KS, TN, A, UT, and CA and countries of Belgium and Germany, Spain, Italy, Greece, France, UK, and Switzerland.
- Recall number
- Z-0824-2011
- Recall status
- Terminated
- CDRH record ID
- 95899
- FDA classified
- 2010-12-28
- Recall event number
- 57341