Open original FDA record ↗Recall record
- FDA identifier
- Z-0824-2013
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Feb 14, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
VS2+ screen may lock up and stop monitoring during patient use
Source facts
- Action
- Philips issued an Urgent Medical Device Correction letter on January 14, 2013. The field action will take place in two phases: Phase 1 Philips sent customer an Urgent Medical Device Correction Notification/Field Safety Notice. This notification explains the issues and gives the customers instructions on actions to take while they await the hardware correction. Phase 2 Once the hardware correction Philips will contact customers and perform the correction free of charge. Until the correction is available. Affected products may continue to be used while waiting for the correction from Philips if the unit is not used in NBP interval mode. A Philips Healthcare representative will contact you to arrange for the correction of all affected devices. If you need any further information or support concerning this issue, please contact the Philips Customer Care Center at 1-800-722-9377. On 3/14/13, firm stated Initially the fix would be a hardware upgrade but have since changed that to a software update. Customers within the US will be sent a new Field Safety Notice as well as a USB software update package and will be required to notify us upon the upgrade of their affected units. They will be performing the updates themselves
- Initiated
- 2013-01-14
- Root cause
- Software design
- Terminated
- 2020-08-27
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: AUSTRALIA, AUSTRIA, BELGIUM,CANADA, CHILE, DENMARK, FRANCE GERMANY, ICELAND, INDIA, IRELAND,ISRAEL,ITALY, JORDAN, KENYA KUWAIT, LEBANON, MALAYSIA, NAMIBIA, NETHERLANDS, NEW ZEALAND NORWAY,PALESTINE, POLAND,PORTUGAL, SAUDI ARABIA, SINGAPORE SLOVENIA, SOUTH AFRICA, SOUTH KOREA,SPAIN,SWEDEN, SWITZERLAND THAILAND, TRINIDAD AND TOBAGO, UNITED ARAB EMIRATES UNITED KINGDOM
- Recall number
- Z-0824-2013
- Recall status
- Terminated
- CDRH record ID
- 115317
- FDA classified
- 2013-02-14
- Recall event number
- 64070