Skip to content

Public FDA record

Z-0824-2018: Camera System, Television, Surgical without audio Intended for transmission and documentation of surgica

Recall record Z-0824-2018 for Trumpf Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0824-2018
Company in FDA record
Trumpf Medical Systems, Inc.
FDA record creation date
Mar 1, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.

Source facts

Action
Trumpf Medical Systems sent an Urgent Medical Device Correction letter dated December 15, 2017. Instructions include to complete and return the Confirmation of Receipt, arrange for the Trumpf Medical Engineer or an authorized Service Engineer to upgrade the affected device, and ensure that all users of the device are aware of the recall. For further questions, please call (843) 329-0543
Initiated
2017-12-15
Root cause
Other
Terminated
2020-04-23
Distribution
Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.
Recall number
Z-0824-2018
Recall status
Terminated
CDRH record ID
162052
FDA record created
2018-03-01
Recall event number
79336

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗