Open original FDA record ↗Recall record
- FDA identifier
- Z-0824-2018
- Company in FDA record
- Trumpf Medical Systems, Inc.
- FDA record creation date
- Mar 1, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.
Source facts
- Action
- Trumpf Medical Systems sent an Urgent Medical Device Correction letter dated December 15, 2017. Instructions include to complete and return the Confirmation of Receipt, arrange for the Trumpf Medical Engineer or an authorized Service Engineer to upgrade the affected device, and ensure that all users of the device are aware of the recall. For further questions, please call (843) 329-0543
- Initiated
- 2017-12-15
- Root cause
- Other
- Terminated
- 2020-04-23
- Distribution
- Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.
- Recall number
- Z-0824-2018
- Recall status
- Terminated
- CDRH record ID
- 162052
- FDA record created
- 2018-03-01
- Recall event number
- 79336