Z-0824-2020: TruDi NAV Suction, TDNS000Z, 0 Degree Tip Angle, (01)10705031245877

Recall record Z-0824-2020 for Acclarent, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0824-2020
Company in FDA record
Acclarent, Inc.
FDA record creation date
Jan 19, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.

Source facts

Action
On 3/6/19, affected customers were contacted via Phone concerning the recall. On 3/12/19, the firm mailed a "Urgent Medical Device Recall" letter to affected customer via FedEx overnight. In addition to informing the customers about the recall, the letter asked customers to take the following actions: 1. Evaluate your current inventory of Acclarent TruDi Nav Suction Instruments with the identified lot numbers listed above. If you have inventory of the specific lots, DO NOT USE. Remove and return products immediately. 2. To return the products, complete BOTH customer acknowledgment and product return sections of the attached Business Reply Form (BRF)). A local Acclarent sales representative will contact you to retrieve the recalled products. Use the enclosed prepaid shipping label and attach a copy of the completed BRF with your product return. 3. Complete the BRF within three (3) business days of receipt even if you no longer have inventory of the identified products. Upon receipt of the completed BRF and your product return, your account will be credited. Incomplete BRFs cannot be processed. 4. Ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully. 5. Maintain a copy of this communication where the inventory of the lots identified in this letter is located until all products are returned. 6. If you have product complaints or adverse events to report regarding the use of Acclarent products, please inform Acclarent by calling (650) 687-5888 7. For questions related to this issue and product recall, please contact your Acclarent sales representative or call (650) 687-5888
Initiated
2019-03-06
Root cause
Device Design
Distribution
US: WI, IL, NE, CA, OH, NY OUS: None
Recall number
Z-0824-2020
Recall status
Terminated
CDRH record ID
178238
FDA record created
2020-01-19
Recall event number
84398

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗