Z-0824-2025: ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL

Recall record Z-0824-2025 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0824-2025
Company in FDA record
Boston Scientific Corporation
FDA classification date
Jan 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Source facts

Action
The firm sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2024. The letter describes the product, problem and actions to be taken. The customers were instructed to promptly identify patients, if device enters Safety Mode, perform emergent replacement for patients who are at risk for harm, during replacement of a device in Safety Mode pacing inhibition should be anticipated during electrocautery. General prophylactic replacement is not recommended; schedule device replacement promptly when the longevity remaining reaches four (4) years or if the longevity remaining is less than 4 years, perform system follow-ups in accordance with the instructions for use and update patients records with letter to maintain awareness for the remaining service life of the device. If you have any questions regarding this information or would like to report a clinical event, contact your Boston Scientific representative or Technical Services. Supplement March 26, 2025, High-Level Overview of Recall Strategy: on 03/26/2025, BSC notified only those implanting and following clinicians responsible for the newly identified active advisory devices. The letter sent to these clinicians will be the original December 12, 2024, Product Advisory letter, however, the March 2025 communication included a cover letter stating that new devices were identified that should have been initially categorized in the Advisory population.
Initiated
2024-12-12
Root cause
Process design
Distribution
US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.
Recall number
Z-0824-2025
Recall status
Open, Classified
CDRH record ID
211665
FDA classified
2025-01-10
Recall event number
95969

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗