Skip to content

Public FDA record

Z-0825-2015: Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular

Recall record Z-0825-2015 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0825-2015
Company in FDA record
DePuy Orthopaedics, Inc.
FDA classification date
Dec 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the

Source facts

Action
On 11/7/2014, URGENT INFORMATION - RECALL NOTICE notifications were sent to the affected distributors, hospitals, and doctors with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospitals and doctors were notified via sales consultant visit. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT). For questions about device recall information provided, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).
Initiated
2014-11-07
Root cause
Mixed-up of materials/components
Terminated
2015-03-24
Distribution
Distributed in the states of WV, AZ, and AL, and the countries of Germany, and Finland.
Recall number
Z-0825-2015
Recall status
Terminated
CDRH record ID
132067
FDA classified
2014-12-17
Recall event number
69984

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗