Z-0825-2022: DigniCap Delta Scalp Cooling System, Product Number: 900-1001, Labeling: The DigniCap Scalp Cooling Syste

Recall record Z-0825-2022 for Dignitana, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0825-2022
Company in FDA record
Dignitana, Inc.
FDA record creation date
Mar 24, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instructions for Use (IFUs) updated to include use of a headband is now recommended for all patients and instructions for the application of the headband. Additionally, IFU updated to include warnings for risk of condensation build up depending on environmental conditions with instructions.

Source facts

Action
On 11/05/2021, emails with subject "RE: DigniCap Delta Scalp Cooling System - Notification of Safety Updates to Instructions for Use." were sent to customers. On 02/25/2022, "URGENT: MEDICAL DEVICE CORRECTION" notices were sent to customers who were informed of the following: All patients should use a headband to prevent direct skin contact with the inner cooling cap. Depending on environmental conditions (e.g., high humidity and/or high ambient temperature) users may experience condensation that could potentially accumulate on certain components of the device such as the connectors and Therapy Hose. This condensation may transfer to the area around the device potentially creating a "Wet floor" condition which could lead to injury from slipping and/or falling. Wipe down the device and any wet components, during or after treatment, as needed at any point in time to avoid hazardous situations. ACTIONS TO BE TAKEN BY THE CUSTOMER/USER: 1. Immediately start using updated instructions amended in this letter. This is a long term and permanent change. 2. Retrain staff to ensure that the headband is used correctly for every treatment. 3. Retrain staff that any condensation on the device should be wiped down before, during and after use 4. Amend the appropriate section of the DigniCap User Manual with the corrections attached to this letter. An electronic copy of the updated User Manual will be emailed to you by March 15, 2022. 5. For orders placed after March 15, 2022, labeling on Affected Guides detailed below will include corrected IFUs. Complete and return the acknowledgment and response form to: Email: qualityaffairs@dignitana.com Postal mail: Dignitana, 10925 Estate Lane W-185, Dallas, TX 75238 USA Dignitana, Traktorgrnden 3, 226 60 Lund, Sweden If you have questions regarding any of the above information, please email support@dignicap.com or contact Dignitana Monday - Friday, 9:00am to 4:00pm, CST +1 877-350-2150.
Initiated
2021-11-05
Root cause
Labeling design
Terminated
2024-10-29
Distribution
US Distribution to states of: LA, NC, NJ, FL, WI, CA, NY, KS, GA, SC, CT and KY.
Recall number
Z-0825-2022
Recall status
Terminated
CDRH record ID
190389
FDA record created
2022-03-24
Recall event number
89056

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗