Open original FDA record ↗Recall record
- FDA identifier
- Z-0827-2016
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Feb 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
Source facts
- Action
- Draeger Medical, Inc. will send an Urgent Medical Device Recall Letter to consignees in a method that confirms delivery. For consignees that have received new devices with firmware version 1.50 (127 devices) or through device updates (644 devices; reference Recall Z-0436-2016), those devices will be downgraded to firmware version 1.49 and the batteries replaced. Batteries will continue to be replaced for all consignees every six months until a final solution is available.
- Initiated
- 2016-02-04
- Root cause
- Software design
- Terminated
- 2020-04-01
- Distribution
- US Nationwide Distribution including Puerto Rico.
- Recall number
- Z-0827-2016
- Recall status
- Terminated
- CDRH record ID
- 143615
- FDA classified
- 2016-02-25
- Recall event number
- 73241