Open original FDA record ↗Recall record
- FDA identifier
- Z-0827-2022
- Company in FDA record
- Mevion Medical Systems, Inc.
- FDA record creation date
- Mar 25, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Source facts
- Action
- Mevion has voluntarily recalled Hand Pendants for S250 and S250i Proton Therapy Systems. All distributed Hand Pendants have been inspected for the identified defect and those that failed inspection have been removed from service following technical service bulletin TSB0048. Subsequent Urgent Medical Device Recall Notice dated 02/11/2022 was hand delivered by on-site field service engineers. Customers are instructed to complete and return the acknowledgement receipt. Please contact Mevion Customer service (robyn.walker@mevion.com) with any questions or concerns.
- Initiated
- 2022-02-01
- Root cause
- Component design/selection
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of FL, MO, NJ, OH, OK, UT, DC and the country of Netherlands.
- Recall number
- Z-0827-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192369
- FDA record created
- 2022-03-25
- Recall event number
- 89758