Skip to content

Public FDA record

Z-0828-2020: Medicrea Pass LP Blunt K-wire 01.6mm x 500mm-Nitinol Product reference: A06010140

Recall record Z-0828-2020 for Medicrea International. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0828-2020
Company in FDA record
Medicrea International
FDA record creation date
Jan 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Conical Cannulated Screws may not accept 1.6mm K-Wire, it cannot enter the minimum tolerance axis threaded

Source facts

Action
Medicrea International issued letter dated 9/12/19 to US Distributor stating reason for recall, health risk and action to take: Immediately check your internal inventory and quarantine all subject devices pending return to MEDICREA 2. Circulate this Field Safety Notice internally to all interested/affected parties. 3. Inform MEDICREA if any of the subject devices have been distributed to other organisations. Please inform MEDICREA of any adverse events concerning the use of the subject devices 2. Please comply with any local laws or regulations concerning the notification of adverseevents to your National Competent Authority. 3. Complete the attached customer response form. It may be that you no longer have any physical inventory on site. Contact: Karine TROGNEUX, MEDICREA INTERNATIONAL Regulatory Affairs Manager, Direct : +33(0)4.69.85.95.39, Standard : +33 (0)4.72.01.87.87. Fax - N +33 (0)04 72 01 87 88, Mail : ktrogneux@medicrea.com
Initiated
2019-09-16
Root cause
Process control
Distribution
Nationwide
Recall number
Z-0828-2020
Recall status
Open, Classified
CDRH record ID
178702
FDA record created
2020-01-24
Recall event number
84595

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.