Reason stated in the source

Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.

Source facts

Action
A Customer Bulletin dated 11/16/21 was sent to customers. Analyzer Return Procedure Our records indicate that you have one of the affected cobas Liat analyzers listed on the previous page. Because of the issue, Roche will send you a new cobas Liat analyzer. Please return your affected analyzer immediately to Roche for investigation by following the steps below: 1. A representative from the Roche Support Network Customer Support Center will contact you to initiate analyzer replacement and confirm shipping information. 2. When you receive the new cobas Liat analyzer from Roche, remove the new analyzer and FedEx shipping label from the box. 3. Place the affected analyzer in the box and seal it using packing tape. 4. Affix the FedEx postage paid shipping label to the outside of the box. 5. Take the box containing the affected analyzer to an appropriate FedEx location for return shipping at no charge to you. Actions Required " Follow the steps in the Analyzer Return Procedure section above. " File this Customer Bulletin for future reference. Please contact the Roche Support Network Customer Support Center at 1-800-800-5973 if you have questions about the information in this Customer Bulletin.
Initiated
2021-11-16
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.
Recall number
Z-0828-2022
Recall status
Open, Classified
CDRH record ID
192098
FDA record created
2022-03-28
Recall event number
89643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗