Skip to content

Public FDA record

Z-0829-03: C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit

Recall record Z-0829-03 for Steris Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0829-03
Company in FDA record
Steris Corp
FDA classification date
May 15, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process.

Source facts

Action
The firm notified their customers by telephone and letter on 4/23/2003.
Initiated
2003-04-23
Root cause
Other
Terminated
2004-02-05
Distribution
The product was shipped to end-users located nationwide and worldwide.
Recall number
Z-0829-03
Recall status
Terminated
CDRH record ID
27177
FDA classified
2003-05-15
Recall event number
26151

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗