While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Source facts
Action
Fresenius Medical Care sent an Urgent Medical Device Correction letter dated April 24, 2017 to affected customers. the letter identified the affected product, problem and actions to be taken. For questions contact Technical Services at 1-800-227-2572.
Initiated
2017-10-05
Root cause
Software Design Change
Terminated
2021-07-28
Distribution
US Nationwide Distribution
Recall number
Z-0829-2018
Recall status
Terminated
CDRH record ID
161733
FDA classified
2018-02-20
Recall event number
79239