Skip to content

Public FDA record

Z-0829-2022: Liko Universal Twinbar 670 QRH, an accessory that can be used together with most Liko lifts, model number

Recall record Z-0829-2022 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0829-2022
Company in FDA record
Baxter Healthcare Corporation
FDA record creation date
Mar 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect center bolt was used in the assembly of the product which could result in a patient fall during use.

Source facts

Action
The recalling firm issued letters to U.S. consignees on 2/21/2022 via UPS ground. The letter provided a description of the problem with the background which included photographs and views of the issue. The potential risk was explained and actions to be taken by the customer were provided, which included to identify whether they have an impacted product and to return the enclosed response form within two weeks. Upon receipt of the response form, Hillrom will contact the consignee to arrange to inspect the Twinbar and if the short bolts have been used, the Twinbar will be corrected. The consignee was instructed not to use the Twinbar together with a Liko Scale until an inspection has been completed. Distributors were instructed to share the Medical Device Correction with their end users and complete the attached response form for return within two weeks. Hillrom will provide replacements for distributors to facilitate exchange among end users. A 3rd party recall service will be issuing notifications to the OUS consignees.
Initiated
2022-02-21
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IN, KS, KY, LA, MA, MD, MI, MN, MO, ND, NE, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. There was government distribution but no military distribution. The countries of Austria, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, Ireland, Japan, Jordan, Latvia, Netherlands, Portugal, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Kingdom, and Uruguay.
Recall number
Z-0829-2022
Recall status
Open, Classified
CDRH record ID
192033
FDA record created
2022-03-28
Recall event number
89598

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗