Lot 0001491 has a manufacturing defect in the sensor spacing. Sensors have been placed 5 cm apart instead of 15 cm apart. Medtronic has determined that patient safety is not compromised. Clinicians however should review and reassess the results of any patient studies for which catheters from lot 0001491 were used.
Source facts
Action
An Urgent Recall letter was issued to all affected customers 1/25/2006. This letter described the issue and listed what actions customers were to take to return affected product.
Initiated
2006-01-25
Root cause
Other
Terminated
2006-12-21
Distribution
within US only: MI, PA, TX
Recall number
Z-0830-06
Recall status
Terminated
CDRH record ID
44031
FDA classified
2006-05-03
Recall event number
34535
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.