Open original FDA record ↗Recall record
- FDA identifier
- Z-0830-2021
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Jan 11, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.
Source facts
- Action
- On 11/23/2020, STERIS issued Urgent Device Correction notices to customers via courier service. Instructing customers to inspect and destroy any remaining affected products in their inventory. STERIS will provide replacement product at no cost to Customers who have product remaining in stock.
- Initiated
- 2020-11-19
- Root cause
- Device Design
- Terminated
- 2023-10-25
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Canada, China, Germany, Japan, Norway, Poland, Qatar, Romania, Sweden, Thailand.
- Recall number
- Z-0830-2021
- Recall status
- Terminated
- CDRH record ID
- 184816
- FDA record created
- 2021-01-11
- Recall event number
- 86906