Skip to content

Public FDA record

Z-0830-2022: Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used

Recall record Z-0830-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0830-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Mar 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may lead to repeat and follow-up testing.

Source facts

Action
The firm, Siemens Healthineers, sent "URGENT MEDICAL DEVICE CORRECTION" (VC-22-01.A.US) on 01/28/2022 via Fed Ex to all customers to inform them of the issue and instruct them to discontinue use and discard the affected lots. Complete and return the Field Correction Effectiveness Check Form via email at uscctsfcaecfax.team@siemens-healthineers.com or fax to: (312) 275-7795. At the present time, Siemens has no replacement lots to offer customers. Siemens suggests using an alternate methodology for magnesium until a replacement lot is available. If you received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Remote Services Center or your local Siemens Technical Support Representative. Replacement kits will be provided when a new lot is available. If you have any questions, please contact your Siemens Healthineers Remote Services Center or your local Siemens Healthineers Technical Support Representative.
Initiated
2022-01-28
Root cause
Under Investigation by firm
Distribution
Worldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY) and OUS countries of: Australia, Canada and Netherlands.
Recall number
Z-0830-2022
Recall status
Open, Classified
CDRH record ID
192233
FDA record created
2022-03-28
Recall event number
89697

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗