Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
Source facts
Action
Integra sent Urgent- Voluntary Medical Device Recall Letters/Recall Acknowledgement and Return Form dated 11/7/2014 to customers and distributors via traceable courier service.
Initiated
2014-11-07
Root cause
Device Design
Terminated
2015-08-11
Distribution
Nationwide, Foreign
Recall number
Z-0831-2015
Recall status
Terminated
CDRH record ID
131571
FDA classified
2014-12-18
Recall event number
69803
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.