Skip to content

Public FDA record

Z-0831-2025: nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which

Recall record Z-0831-2025 for Nuwellis Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0831-2025
Company in FDA record
Nuwellis Inc
FDA classification date
Jan 16, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.

Source facts

Action
Nuwellis issued an UREGNT: MEDICAL DEVICE RECALL notice to its consignees on 12/11/2024 via email. The notice explained the issue, risk to health, and requested the following actions be taken: Immediately discontinue use of affected lots. A Nuwellis representative will be reaching out to arrange return or disposal of the affected product to Nuwellis. For more detailed discussion regarding this product return, you may contact the Quality Team at Nuwellis: Steve Sandoval, Director of Engineering and Quality, Email: Steve.Sandoval@nuwellis.com, Office Phone: +1 952.563.7040
Initiated
2024-12-11
Root cause
Under Investigation by firm
Distribution
US Nationwide Distribution and the country of Hong Kong
Recall number
Z-0831-2025
Recall status
Completed
CDRH record ID
211717
FDA classified
2025-01-16
Recall event number
95991

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗