The firm, Philips Healthcare, sent an "URGENT - Medical Device Correction" letter dated December 9, 2013 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following to avoid the occurrence of these specific ring artifacts:
1. Disable Automatic Scan Time for all Exam Cards
2. Leave Automatic Scan Time on during planning and prior to clicking Go
to initiate the scan, click the acquisition series of the scan and disable
Automatic Scan Time, then proceed with the scan.
3. Disable Edit Before Final Recon (Reconstruction) for all Exam Cards
If a ring artifact is present, clinical judgment should be used by the
customer/user to determine if the images acquired in the dataset contain enough information to make a diagnosis.
Philips will be releasing a Field Change Order (FCO) to correct the issue with software version release 4.1.1.19261. In addition, a Philips Field Service Engineer will contact you to schedule the software installation at your site.
If you need any further information or support concerning this issue, please
contact your local Philips representative or local Philips Healthcare office. For
North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
Initiated
2013-12-09
Root cause
Software design
Terminated
2014-05-29
Distribution
Worldwide Distribution: US (nationwide) including states of: FL, IN, NY, OR and PA; and internationally to: Australia and Israel.
Recall number
Z-0832-2014
Recall status
Terminated
CDRH record ID
125215
FDA classified
2014-01-27
Recall event number
67306