Skip to content

Public FDA record

Z-0832-2015: HemosIL PT-Fibrinogen HS PLUS, IVD --- Instrumentation Laboratory IL Company. A high sensitivity thrombo

Recall record Z-0832-2015 for Instrumentation Laboratory Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0832-2015
Company in FDA record
Instrumentation Laboratory Co.
FDA classification date
Dec 19, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some vials of HemosIL PT-Fibrinogen HS PLUS, Part No. 0008469810 (various lots manufactured prior to June 2014) exhibit a yellow/brownish color after reconstitution (yellow color or darker) instead of the typical white to off-white color. Vials exhibiting this yellow/brownish color were observed to cause prolonged prothrombin (PT) clotting times.

Source facts

Action
The firm, Instrumentation Laboratory, sent an "URGENT MEDICAL DEVICE CORRECTION" letter (dated 11/21/2014) to US and Canadian customers via USPS on 11/21/2014 for all in-date lots of HemosiL PT-Fibrinogen HS PLUS, Part No. 0008469810, manufactured prior to June 2014. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: if they observe any yellow/brownish color after reconstitution, they are not to use the affected vial; discard the vial and reconstitute another vial; only use a vial if the reconstituted color is the typical white to off-white color; run quality controls at least once every 8 hour shift in accordance with GLP and insert instruction sheet; share this information with your laboratory staff and follow initial procedures; forward this notification to all affected locations within your facility; retain a copy of this notification for your records and complete and return the tracking record form within 10 business days of receiving notification even if you do not use these reagents via fax at 781-861-4207 or email to: ra-usa@ilww.com. On 11/21/2014, e-mail notifications were sent to the IL/Werfen Country Affiliates and International Group (who manage the local country dealers and distributors) to prepare and issue Field Safety Notices in the respective languages to their customers. For technical questions, please contact the IL Technical Support Center at 1-800-678-0710, Option # 2 (ACL Coagulation Products). For additional questions, please contact your IL Customer Service Representative at 1-800-955-9525, Option # 2.
Initiated
2014-11-21
Root cause
Process design
Terminated
2016-01-26
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: ALBANIA, ALGERIA, ANDORRA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BELARUS, BRAZIL, BRUNEI, BULGARIA, CHILE, CHINA, COLUMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, FRANCE, GABON, GEORGIA, GERMANY, GREECE, GUATEMALA, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KENYA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA, MEXICO, NEW ZEALAND, PAKISTAN, PANAMA, PARAGUAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TURKEY, UAE, UGANDA, UKRAINE, UNITED KINGDOM, AND URUGUAY.
Recall number
Z-0832-2015
Recall status
Terminated
CDRH record ID
131733
FDA classified
2014-12-19
Recall event number
69824

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗