Skip to content

Public FDA record

Z-0832-2021: Naturalyte Liquid Acid concentrate bottles, Product Code 08-2201-5 - Product Usage: used in conjunction w

Recall record Z-0832-2021 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0832-2021
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Jan 11, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 11/23/20 was sent to customers. Fresenius Medical Care Renal Therapies Group, LLC (FMCRTG) previously issued Urgent Medical Device Alerts on November 7, 2020 and November 14, 2020 alerting you to issues pertaining to the sodium level in the Naturalyte Liquid Acid concentrate and Citrasate Liquid Acid concentrate products impacting 105 lots, manufactured at the Oregon, Ohio facility between September 9, 2020 and October 29, 2020. After further testing, FMCRTG is now voluntarily recalling 28 of the 105 lots for one or both of the following reasons: " The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. " Calcium concentration is below manufacturing specification. The 28 lots were released for distribution between September 30, 2020 and November 20, 2020. Please check your stock and discontinue use and/or distribution of the products. If you have any listed lot numbers, place the product in a secure, segregated area. Please contact Stericycle at 1-866-912-9552 for instructions on how to return the product and receive credit This is EVENT# 5857. o If returning single bottles, please place affected product in a box for pick up. o In order to receive credit for your impacted product, you must call the Stericycle phone number above and provide quantities for return. Additional Information: If urgent replacement of product is needed, please call FMCRTG Customer Service at 1-800-323-5188. Otherwise, please add any additional product request to your next scheduled delivery. We apologize for any inconvenience. Please complete and return the enclosed Reply Form at your earliest convenience and contact our Customer Services team at 1-800-323-5188 with any additional questions. Please contact our Medical Information team at 1-855-616-2309 to discuss any medical or scie
Initiated
2020-11-23
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0832-2021
Recall status
Open, Classified
CDRH record ID
184863
FDA classified
2021-01-11
Recall event number
86939

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗