Skip to content

Public FDA record

Z-0833-03: Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instrument

Recall record Z-0833-03 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0833-03
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
May 15, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change.

Source facts

Action
Baxter sent recall letters dated 3/20/03 to the Medical Director and Hemodialysis Chief Technicians via first class mail on 3/20/03. The letters informed the accounts of the potential of running a treatment without bicarbonate due to an undetected bicarbonate pump stall following a sodium change, and provided instructions to change the B-A-alarm limits on the System 1000 Delivery Systems and confirm proper functionality after completing these changes. These changes will prevent undetected acid-only delivery in the event of a sodium level change
Initiated
2003-03-20
Root cause
Other
Terminated
2004-01-23
Distribution
The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Belgium, Bolivia, Brazil, Canada, Chile, China, Ecuador, Egypt, El Salvador, Guatemala, Honduras, Hong Kong, India, Mexico, Philippines, Singapore, South Korea, Thailand, Turkey, United Kingdom, Venezuela and Vietnam.
Recall number
Z-0833-03
Recall status
Terminated
CDRH record ID
26530
FDA classified
2003-05-15
Recall event number
25818

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗