Product fails to comply with the federal performance standard for cabinet x-ray systems (Title 21 CFR 1020.40(c)(1)(i)). Specifically, the x-ray emission limit of 0.5 milliroentgen in one hour could be exceed. Highest emission measured was 3 mR/hr.
Source facts
Action
GE Healthcare will notify purchasers of the problem, the affected product, instructions for users, and specify the maximum emission rate measured. GE Healthcare will bring defective systems into compliance free of charge. This will be done via Field Modification Instructions IFMI) 11101 which is scheduled for deployment on March 1, 2011 with an estimated completion date of July 1, 2011.
Initiated
2011-01-31
Root cause
Radiation Control for Health and Safety Act
Terminated
2012-09-06
Distribution
Worldwide Distribution
Recall number
Z-0833-2012
Recall status
Terminated
CDRH record ID
97538
FDA classified
2012-02-17
Recall event number
57842