Open original FDA record ↗Recall record
- FDA identifier
- Z-0833-2013
- Company in FDA record
- Ventana Medical Systems Inc
- FDA classification date
- Aug 6, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining platform or the DISCOVERY ULTRA automated staining platform. This type of overflow event can lead to migration of fluid across tubing and wiring within the instrument cabinet resulting in contact with electrical circuitry. This may create shorting conditions and u
Source facts
- Action
- The firm, Roche sent a "Technical Bulletin" and an letter dated October 3, 2012 to its customers. The letter and bulletin described the product, problem and actions to be taken. The customers were informed that Ventana is aggressively working on solutions to this issue. Until mitigations to the overflow condition are available for implementation; inform all personnel in your organization that work with the BenchMark ULTRA or DISCOVERY ULTRA instrument about the potential for fluid overflow to occur during decontamination cycles; follow the instructions provided in the letter to avoid the documented failure mode and follow the instructions provided in the technical bulletin. Contact your local support center with any questions or concerns or call 520-877-7191.
- Initiated
- 2012-09-28
- Root cause
- Device Design
- Terminated
- 2016-02-24
- Distribution
- Worldwide distribution: USA (nationwide) and country of: Germany.
- Recall number
- Z-0833-2013
- Recall status
- Terminated
- CDRH record ID
- 115665
- FDA classified
- 2013-08-06
- Recall event number
- 64244