Open original FDA record ↗Recall record
- FDA identifier
- Z-0833-2015
- Company in FDA record
- Ge Healthcare It
- FDA classification date
- Dec 19, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.
Source facts
- Action
- General Electric sent an Urgent Medical Device Correction letter dated April 19, 2013, to all affected customers. The letter instructs customers to ensure that Enterprise Archive partitions have enough free space available, in particular the cache disk should have at least 10% free space available. This could be facilitated by setting appropriate disk limits and watermarks on the individual libraries in Enterprise Archive. In case the disks are not being purged to the configured watermarks and the cache disk is almost full, customers are requested to contact their local GE service person. A permanent product correction will be provided to all customers under FMI85209R. Customers with questions were instructed to contact their local Sales/Service Representative. Also the Remote Online Center (ROC) Customer Care can be reached at 1-800-437-1171.
- Initiated
- 2013-04-19
- Root cause
- Software design
- Terminated
- 2016-06-10
- Distribution
- Nationwide Distribution including AZ, OH, WA.
- Recall number
- Z-0833-2015
- Recall status
- Terminated
- CDRH record ID
- 117004
- FDA classified
- 2014-12-19
- Recall event number
- 64760