Z-0833-2015: Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data.

Recall record Z-0833-2015 for Ge Healthcare It. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0833-2015
Company in FDA record
Ge Healthcare It
FDA classification date
Dec 19, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.

Source facts

Action
General Electric sent an Urgent Medical Device Correction letter dated April 19, 2013, to all affected customers. The letter instructs customers to ensure that Enterprise Archive partitions have enough free space available, in particular the cache disk should have at least 10% free space available. This could be facilitated by setting appropriate disk limits and watermarks on the individual libraries in Enterprise Archive. In case the disks are not being purged to the configured watermarks and the cache disk is almost full, customers are requested to contact their local GE service person. A permanent product correction will be provided to all customers under FMI85209R. Customers with questions were instructed to contact their local Sales/Service Representative. Also the Remote Online Center (ROC) Customer Care can be reached at 1-800-437-1171.
Initiated
2013-04-19
Root cause
Software design
Terminated
2016-06-10
Distribution
Nationwide Distribution including AZ, OH, WA.
Recall number
Z-0833-2015
Recall status
Terminated
CDRH record ID
117004
FDA classified
2014-12-19
Recall event number
64760

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗