Z-0834-2020: Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Recall record Z-0834-2020 for Polymer Technology Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0834-2020
Company in FDA record
Polymer Technology Systems, Inc.
FDA record creation date
Jan 21, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Source facts

Action
Urgent Medical Device Correction notification letters dated 11/15/19 were sent to customers. Actions to be taken by the Customer/Distributor/User: 1. Immediately check your inventories and apply the sticker/labels as directed. If you are an end user: Please adhere one label to the outside of the battery door compartment of your analyzer and apply the other label to the analyzer user guide. If you are a distributor: Please remove the tape liner on the back of the large sticker leaving both small labels intact for the end user to apply. Apply the large sticker to the top panel of the analyzer box, and press firmly. 2. If you have further distributed this product, immediately identify the users and notify them at once of this product correction. Forward this notice, stickers, and response form to all customers/users who did, or who may have, received, this product. 3. Once you have taken the appropriate action, please fill out the attached response form and return it to PTS Diagnostics via the information below. It is important that you complete and return the response form, even if you do not have affected product on hand. 4. Please contact PTS for additional stickers and a PDF version of this letter and response form, if needed. Thank you for your prompt action, and we apologize for any inconvenience this issue may have caused you and your facility. If you have any questions, please contact Customer Service, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 319-870-5610 or 877-870-5610 (U.S. toll-free), or via email at response@ptsdiagnostics.com.
Initiated
2019-12-02
Root cause
Device Design
Terminated
2020-07-06
Distribution
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.
Recall number
Z-0834-2020
Recall status
Terminated
CDRH record ID
178369
FDA record created
2020-01-21
Recall event number
84274

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗