Z-0834-2022: FORE-SIGHT ELITE Absolute Tissue Oximeter, REF: 01-06-3000, Contents: Monitor 1 EA., Accessory Kit 1 EA.

Recall record Z-0834-2022 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0834-2022
Company in FDA record
Edwards Lifesciences, LLC
FDA record creation date
Mar 30, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

Source facts

Action
Urgent Medical Device Correction communication dated March 01, 2022 was sent via FedEx 2nd Day Advice on action to be taken by user: Customers are to refrain from using the large sensors on certain somatic locations (arms and legs). Cerebral and flank/abdomen locations for the large sensors are not impacted. Instructions for customers: " Review the customer letter for advice on how to use impacted product. " Please follow the instructions included in the enclosed acknowledgement form to complete the acknowledgement process. " Do not return any product. " Distribute this notice within your organization or to any organization where the potentially impacted product has been transferred. " Verify your inventory. " E-mail the completed form to Edwards Technical Support at tech_support@edwards.com, within 15 days from receipt of this notification. Advice on action to be taken by Distributor: Please review this letter and complete the acknowledgement form. Return the acknowledgement form to tech_support@edwards.com within 15 days of receipt of this notification. Please forward this customer communication to any of your customers who have purchased the impacted Edwards product. If you still have product within your control, please do not distribute impacted product to any customers. You can call Edwards Tech Support for instructions for handling product still within your control. Alternatively, you can provide Edwards with a list of your customers who have purchased the impacted product from you and Edwards will communicate directly with your customers to facilitate the correction and acknowledgement process. Please forward your customer list to tech_support@edwards.com. If you have any questions, please contact Edwards Tech Support at 1 (800)-822-9837 If you have questions, contact Edwards Tech Support at 1 (800)-822-9837 from 5:00AM--4:30PM PST.
Initiated
2022-03-01
Root cause
Software in the Use Environment
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, CANARY ISLANDS, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Northern Ireland, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), United Kingdom, Vietnam.
Recall number
Z-0834-2022
Recall status
Open, Classified
CDRH record ID
192119
FDA record created
2022-03-30
Recall event number
89647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗