Skip to content

Public FDA record

Z-0835-06: EL-aCL Anti-Cardiolipin ELISA Kit; an in-vitro diagnostic anticardiolipin immunological test system; 192

Recall record Z-0835-06 for Thera Test Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0835-06
Company in FDA record
Thera Test Laboratories, Inc.
FDA classification date
May 4, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The specimen diluent contained in the kits was contaminated with a fungal growth. Also, a portion of kit lot 09055514 was shipped without the Data Sheet, part #201-204.

Source facts

Action
On 10/26/05, Thera Test telephoned the 3 accounts who received lot 09055514 without the Data Sheet in the kits, and sent them copies of the Data Sheet for insertion into the 6 kits on hand at the consignees. Between 1/4-9/06, Thera Test telephoned the 17 accounts receiving kit lots 09055514 and 12055603, and sent them follow-up letters dated 1/9/06 with replacement bottles of Specimen Diluent. The accounts were advised of the fungal contamination found in some bottles of the diluent shipped with the kit lots, and were requested to destroy any remaining bottles of the Specimen Diluent in their kits and replace them with the bottles of diluent sent with the letter.
Initiated
2006-01-11
Root cause
Other
Terminated
2006-05-03
Distribution
California, Connecticut, Maryland, Pennsylvania, Ohio, New Jersey and Florida.
Recall number
Z-0835-06
Recall status
Terminated
CDRH record ID
44985
FDA classified
2006-05-04
Recall event number
34945

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗