Z-0835-2025: Brand Name: Alcon Product Name: Monarch III D Delivery Cartridge Model/Catalog Number: 8065977763 Produc

Recall record Z-0835-2025 for Alcon Research LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0835-2025
Company in FDA record
Alcon Research LLC
FDA classification date
Jan 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery.

Source facts

Action
On 11/13/2024, Alcon Laboratories, Inc. issued a "Urgent: Medical Device Field Correction" to affected consignees. Alcon asked consignees to take the following actions: 1. See Table 1 for list of affected of Monarch III D delivery cartridges distributed to your facility. 2. Review your Monarch III D delivery cartridge inventory to determine if you have any remaining units from the identified lots. 3. If you have remaining units from the identified lots, please dispose of all impacted lots of Monarch III D delivery cartridges identified in Table 1. Please contact Alcon Customer Service to arrange for replacement of Monarch III D delivery cartridges. 4. Please confirm quantity destroyed by including on response form, even if you have zero units remaining in inventory. 5. Forward this notification to professionals within your organization who may be using the identified of Monarch III D delivery cartridges, and to all departments or organizations to which identified lots of Monarch III D delivery cartridges lots may have been transferred. 6. Respond to Alcon indicating your understanding of the included instructions by completing our online response form at https://qrco.de/Monarch-Response or by completing the attached Response Form and returning to Alcon via email or fax. 7. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email (msus.safety@alcon.com) or by phone (1-800-757-9780) Monday through Friday, 8:00 AM to 5:00 PM, Central Time. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm.
Initiated
2024-11-13
Root cause
Process design
Distribution
US: AK AL AR AZ CA CO CT DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY OUS: Russia, Egypt
Recall number
Z-0835-2025
Recall status
Open, Classified
CDRH record ID
211494
FDA classified
2025-01-10
Recall event number
95903

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗