Open original FDA record ↗Recall record
- FDA identifier
- Z-0835-2025
- Company in FDA record
- Alcon Research LLC
- FDA classification date
- Jan 10, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery.
Source facts
- Action
- On 11/13/2024, Alcon Laboratories, Inc. issued a "Urgent: Medical Device Field Correction" to affected consignees. Alcon asked consignees to take the following actions: 1. See Table 1 for list of affected of Monarch III D delivery cartridges distributed to your facility. 2. Review your Monarch III D delivery cartridge inventory to determine if you have any remaining units from the identified lots. 3. If you have remaining units from the identified lots, please dispose of all impacted lots of Monarch III D delivery cartridges identified in Table 1. Please contact Alcon Customer Service to arrange for replacement of Monarch III D delivery cartridges. 4. Please confirm quantity destroyed by including on response form, even if you have zero units remaining in inventory. 5. Forward this notification to professionals within your organization who may be using the identified of Monarch III D delivery cartridges, and to all departments or organizations to which identified lots of Monarch III D delivery cartridges lots may have been transferred. 6. Respond to Alcon indicating your understanding of the included instructions by completing our online response form at https://qrco.de/Monarch-Response or by completing the attached Response Form and returning to Alcon via email or fax. 7. In the event you have experienced adverse events or product quality issues related to this communication, please contact Alcon via web (https://notifeye.alcon.com), by email (msus.safety@alcon.com) or by phone (1-800-757-9780) Monday through Friday, 8:00 AM to 5:00 PM, Central Time. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA MedWatch Adverse Event Reporting program: http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm.
- Initiated
- 2024-11-13
- Root cause
- Process design
- Distribution
- US: AK AL AR AZ CA CO CT DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY OUS: Russia, Egypt
- Recall number
- Z-0835-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211494
- FDA classified
- 2025-01-10
- Recall event number
- 95903