Z-0836-2011: Varian Medical Systems FastPlan software, Versions 5.5 and 5.5.1, Model Number HZ3, manufactured by Varia

Recall record Z-0836-2011 for Varian Medical Systems Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0836-2011
Company in FDA record
Varian Medical Systems Oncology Systems
FDA classification date
Dec 29, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Due to software Issue, an error occurs when performing cone planning using MRI, which may result in the patient receiving a higher radiation dose than expected, potentially to healthy tissue.

Source facts

Action
Varian sent Urgent Medical Device Correction letters on March 23, 2010, by Federal Express to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to run test cases or otherwise determine whether the MR images that they receive have the 'Series Date' parameter populated. Customers should contact their MR system manufacturer to get a confirmation that the 'Series Date' field is populated. Customers were instructed to confirm by visual inspection that the fusion results are loaded appropriately in Cone Planning before starting to create or edit a plan. For questions regarding this recall contact your local Varian Customer Support Distirct or Regional Manager. Phone: USA and Canada - 1-888-VARIANS (888-827-4265) Europe -+41 41 749 8844 E-mail: North America: support-america@varian.com Australia/New Zealand: support-anz@varian.com Europe: support-emea@varian.com South East Asia: seasia.apps.helpdesk@varian.com China/Asia: china.apps.helpdesk@varian.com Japan: Japan.Apps.Helpdesk@varian.com Latin America: soporte.al@varian.com Internet: Oncology Systems customer site - www.myvarian.com Varian Medical System public site - www.varian.com
Initiated
2010-03-23
Root cause
Device Design
Terminated
2012-02-24
Distribution
Worldwide distribution - USA nationwide and the countries of Australia, Brazil, Canada, Chile, Colombia, Georgia, Germany, Hong Kong, India, Ireland, Japan, Macau, Malaysia, Russia, Singapore, Spain, Sri Lanka, and Taiwan
Recall number
Z-0836-2011
Recall status
Terminated
CDRH record ID
92440
FDA classified
2010-12-29
Recall event number
56024

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗