Open original FDA record ↗Recall record
- FDA identifier
- Z-0836-2015
- Company in FDA record
- Ion Beam Applications S.A.
- FDA classification date
- Dec 19, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Source facts
- Action
- The firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters, with the correct values. If you need assistance, contact the IBA Site Staff or the Operations manager at +32 10 475 997 or email: Sophie.dessart@iba-group.com.
- Initiated
- 2014-11-07
- Root cause
- Software design (manufacturing process)
- Terminated
- 2015-06-29
- Distribution
- US Distribution in PA only.
- Recall number
- Z-0836-2015
- Recall status
- Terminated
- CDRH record ID
- 131533
- FDA classified
- 2014-12-19
- Recall event number
- 69781