- FDA identifier
- Z-0836-2020
- Company in FDA record
- Paltop Advanced Dental Solutions Ltd
- FDA record creation date
- Jan 21, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Some of the implants were packaged with a Standard Platform Cover Screw instead of a Narrow Platform Cover Screw. The Standard Platform Cover Screw will not fit in the Narrow Platform implant.
Source facts
- Action
- Urgent: Medical Device Recall notification letters dated 11/18/19 were sent to customers. Please discontinue use. Attached is a Reply Communication form that should be used to let us know whether the product has been used or will be returned. Once we receive that Form, a Customer Service agent will reach out to you with return instructions and provide a label if necessary. Paltop Advanced Dental will replace the implants that are in stock.
- Initiated
- 2019-11-18
- Root cause
- Mixed-up of materials/components
- Terminated
- 2022-12-27
- Distribution
- The products were distributed to the following US states: CA, DC, FL, GA, MA, MD, NJ, NY, SC, TX, and VA.
- Recall number
- Z-0836-2020
- Recall status
- Terminated
- CDRH record ID
- 178257
- FDA record created
- 2020-01-21
- Recall event number
- 84412