Z-0836-2024: Terragene Integron IT26-C Terragene Integron IT26-C with extender Benco Dental Chemical Integrator [572

Recall record Z-0836-2024 for TERRAGENE S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0836-2024
Company in FDA record
TERRAGENE S.A.
FDA classification date
Jan 26, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.

Source facts

Action
On 12/07/2023, the firm notified via email its initial importers/distributor a "Voluntary Recall Notification " that contained a "COMMUNICATION TO CUSTOMER VOLUNTARY RECALL OF PRODUCT/URGENT: MEDICAL DEVICE RECALL" Letter informing them of the discrepancies identified in the IFUs of some products that mentions extended cycles that may mislead users to think that these cycles are part of the FDA-cleared indications for use of the product. Initial Importers/Distributors are instructed to send the following information to Terragene: -Quantity in stock -Quantity distributed -Consignees On or about 12/22/2023, the firm initiated distribution, via email, an "URGENT: MEDICAL DEVICE RECALL" Letter to customers (identified by their Initial Imports/Distributors) informing them of the Recall issue (discrepancy identified in the IFU) provide an updated IFU (version 4) correcting the issue. Customers are instructed to print the updated IFU and attach it to their remaining inventory, and to discard the older IFU versions (version 1 thru 3). For questions or concerns, contact customer.service@terragene.com or www.terragene.com , or +1-844-837-7243 Monday thru Friday 6:00 AM to 3:00 PM, Eastern Time.
Initiated
2023-12-07
Root cause
Labeling False and Misleading
Distribution
U.S. Nationwide distribution in the states of GA, IN, MI, MT, NY, PA, TX and VA.
Recall number
Z-0836-2024
Recall status
Open, Classified
CDRH record ID
204984
FDA classified
2024-01-26
Recall event number
93625

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.