Z-0836-2025: Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of

Recall record Z-0836-2025 for Biocartis Nv. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0836-2025
Company in FDA record
Biocartis Nv
FDA classification date
Jan 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

False-positive MSI-H results generated by the Idylla MSI Test.

Source facts

Action
Biocartis issued an URGENT - Medical Device Correction notice to its consignees on 12/10/2024 via email. The notice provided the description of the problem, its impact, and requested the following actions be taken: 1. Until your Instrument is serviced, please move forward with the option that applies to your situation: o In case you have another Idylla Instrument available that is not affected, use that Instrument to run the Idylla MSI Test and discontinue running the Idylla MSI Test on the listed Instrument(s). o No new Idylla MSI testing should be performed using the affected Idylla Instruments. o For results already obtained using an affected Idylla Instrument, the following apply: o MSS results are VALID and may be reported. o MSI-H (Microsatellite Instability High) results with six or seven positive identified biomarkers (as depicted in the result report) are VALID and may be reported. o MSI-H results other than those with six or seven positive identified biomarkers are INVALID and should NOT be reported or used to manage patient health. Patient samples with these results should be re-tested on an Idylla Instrument that is NOT affected. 2. If you are not the Idylla user, please forward this notice to the Idylla end user or Idylla supervisor and provide Biocartis with the correct contact details for our record keeping. 3. Completion of the Acknowledgement of Receipt (Appendix 2) is required for compliance. Please complete and sign the attached Acknowledgement of Receipt form by December 17, 2024, and send it to vigilance@biocartis.com.
Initiated
2024-12-10
Root cause
Environmental control
Distribution
US: AL, NJ, TX, OH, CA
Recall number
Z-0836-2025
Recall status
Open, Classified
CDRH record ID
211800
FDA classified
2025-01-10
Recall event number
96022

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗