Skip to content

Public FDA record

Z-0837-06: 20F Tri-Funnel Replacement Gastrostomy Tube

Recall record Z-0837-06 for Bard Access Systems, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0837-06
Company in FDA record
Bard Access Systems, Inc
FDA classification date
May 3, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Packaging for some gastrostomy tubes was labeled with incorrect balloon inflation data. An underinflated balloon may cause misplacement, slippage or premature removal of the tube. If an underinflated balloon allows the tube to be placed, or slip into, the stomal tract and not the stomach, there is a risk of infusion of fluids into the peritoneum. This carries the risk of infection.

Source facts

Action
Consignees were notified by letter on 04/06/2006.
Initiated
2006-04-06
Root cause
Error in labeling
Terminated
2009-08-12
Distribution
Nationwide, including 4 VA facilities. No military consignees. Foreign distribution to Canada, Australia
Recall number
Z-0837-06
Recall status
Terminated
CDRH record ID
45405
FDA classified
2006-05-03
Recall event number
35140

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.