Skip to content

Public FDA record

Z-0837-2011: Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon

Recall record Z-0837-2011 for B Braun Medical, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0837-2011
Company in FDA record
B Braun Medical, Inc
FDA classification date
Jan 10, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.

Source facts

Action
B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Initiated
2010-08-24
Root cause
Component design/selection
Terminated
2014-05-29
Distribution
Virginia and New York
Recall number
Z-0837-2011
Recall status
Terminated
CDRH record ID
95391
FDA classified
2011-01-10
Recall event number
57099

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.