Recalled devices were packaged in convenience kits.
Source facts
Action
A customer letter dated February 24, 2022 was issued to the impacted consignee. The letter instructs the consignee to perform a microscopy to confirm an elevated RBC count in a urine sample as the impacted component may shed contaminant particles that are identified as RBCs when used in conjunction with the Sysmex UF-1000i or UX-2000 analyzers.
A reply form should be completed and returned to ROi confirming receipt and understanding of the notification.
End users should retain a copy of the BD notice and post near the analyzer as a visual reminder of the potential issue with the BD product.
Initiated
2022-02-24
Root cause
Material/Component Contamination
Terminated
2022-04-25
Distribution
US Nationwide distribution in the state of Missouri.
Recall number
Z-0837-2022
Recall status
Terminated
CDRH record ID
192280
FDA record created
2022-03-30
Recall event number
89710