Skip to content

Public FDA record

Z-0837-2025: Karl Storz - Endoskope - REF 723400 - Optical Biopsy and Grasping Forceps, CE

Recall record Z-0837-2025 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0837-2025
Company in FDA record
Karl Storz Endoscopy
FDA classification date
Jan 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to failure of manual cleaning validation, biopsy and grasping forceps like sterility assurance.

Source facts

Action
On December 19, 2024, Karl Storz issued a "Urgent: Medical Device Recall" Notification to affected consignees. In addition to informing consignees about the recall, Karl Storz ask consignees to take the following actions: 1. Immediately quarantine and discontinue use of associated part numbers listed. 2. Pass on this Urgent Medical Device Recall Notice to all users of the recalled product and all other persons who need to be aware within your organization. 3. If you have further distributed the recall product, please identify and promptly notify those recipients by providing a copy of the recall notification and response form, or provide KARL STORZ a list of customers who received/may have received the product listed. 4. Return the completed Customer Reply Form by Fax or E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Contact your KARL STORZ representative or call KARL STORZ Customer Support to return affected product. " Call Customer Support at 800-421-0837 Option 1 to set up an RMA. " Or call your KARL STORZ representative to assist with setting up an RMA. 6. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2024-12-19
Root cause
Reprocessing Controls
Distribution
US: AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OR, PA, SC, TX, VA OUS: None
Recall number
Z-0837-2025
Recall status
Open, Classified
CDRH record ID
211635
FDA classified
2025-01-10
Recall event number
95954

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗