Z-0838-2019: PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through

Recall record Z-0838-2019 for Leica Microsystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0838-2019
Company in FDA record
Leica Microsystems, Inc.
FDA record creation date
Feb 15, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.

Source facts

Action
The firm initiated their field notification by email on 09/25/2017. The firm provided details on the issues and requested the user to take the following actions: "If you notice any of the anomalies as described above before being contacted by a Leica Microsystems representative, please contact your Leica representative for relevant repair of your PROVEO 8. You can continue to use the system pending the intervention of the appropriate Leica Microsystems representative. However, please ensure that a functional check is performed and the system is properly balanced prior to surgery as required by and described in chapters 17.1 Checklist before operation and 7.7.1 Balancing the parallelogram of the user manual. During surgery, ensure that no functional switches are activated at the same time as a user profile or mode is changed. During the start-up of the system, ensure that no functional switches are activated until the system boots up."
Initiated
2017-09-25
Root cause
Other
Terminated
2020-01-21
Distribution
Worldwide
Recall number
Z-0838-2019
Recall status
Terminated
CDRH record ID
158862
FDA record created
2019-02-15
Recall event number
78190

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗