Z-0838-2020: Imactis Navigation Kit, REF 10100, Content Sterile Needle Holder Ref. A10010 CE 0459, Sensor Cover Ref. P

Recall record Z-0838-2020 for Imactis. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0838-2020
Company in FDA record
Imactis
FDA record creation date
Jan 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

Source facts

Action
On 7/31/2019, Imactis contacted affected consignees via phone and followed up the phone call with an "Urgent Field Safety Notice" sent via E-Mail. In addition to informing the consignees about the recall , the notice also ask customers to take the following action: 1. Identification and quarantine of the concerned devices, 2. Acknowledgement of receipt of the safety information provided in Annex 1 to be returned to the manufacturer. 3. This safety information must be transmitted to all persons in your facility who may be concerned by the implementation of safety corrective actions or to any department to which the removed products could have been transferred. 4. We recommend that you pay particular attention to the required corrective actions in order to ensure compliance. Action plan: 1. A new sensor cover has been identified for use with the IMACTIS CT-Navigation system. This is reference 610-797 and manufactured by CIVCO. Therefore, the following actions will be implemented: 2. Customer will complete Annex 1 and send to quality@imactis.com before August 9, 2019. 3. Customer will complete Annex 2 referencing the removed lot numbers and return to the address provided. 4. Upon receipt of your removed kits, IMACTIS will send back the new I10100 navigation kits containing the cover under reference 610-797.
Initiated
2019-08-01
Root cause
Material/Component Contamination
Distribution
US: WI & IL
Recall number
Z-0838-2020
Recall status
Open, Classified
CDRH record ID
178379
FDA record created
2020-01-22
Recall event number
84460

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗