Packaging for some gastrostomy tubes was labeled with incorrect balloon inflation data. An underinflated balloon may cause misplacement, slippage or premature removal of the tube. If an underinflated balloon allows the tube to be placed, or slip into, the stomal tract and not the stomach, there is a risk of infusion of fluids into the peritoneum. This carries the risk of infection.
Source facts
Action
Consignees were notified by letter on 04/06/2006.
Initiated
2006-04-06
Root cause
Error in labeling
Terminated
2009-08-12
Distribution
Nationwide, including 4 VA facilities. No military consignees. Foreign distribution to Canada, Australia
Recall number
Z-0839-06
Recall status
Terminated
CDRH record ID
45407
FDA classified
2006-05-03
Recall event number
35140