Several units were manufactured using a power supply that contains a 1 amp fuse where a 5 amp fuse is required. When the unit's power supply draws more that 1A, the fuse will fail causing the light to cease operating. The problem can occur anytime: from the initial turning on (powering up), to anytime during use.
Source facts
Action
Notification sent 04/09/07 via FedEx to distribution centers, sales reps and user facilities. Users asked to send light source to Smith & Nephew GMBH Service Center in Germany for repair (contact Ute Dieling at +49 (0) 7462-208-313).
Initiated
2007-04-09
Root cause
Nonconforming Material/Component
Terminated
2008-09-02
Distribution
Product sent to distributors in Argentina, Austria, Chile, Denmark, Dubai, France, Greece, Hong Kong, India, Italy, Japan, Korea, Malaysia, Norway, South Africa, Spain, Sweden and United Kingdom. No US distribution.
Recall number
Z-0840-2007
Recall status
Terminated
CDRH record ID
51944
FDA classified
2007-06-07
Recall event number
37849