Software issues including:
Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan;
Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card;
Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed;
Software errors that may result in partial set of images or inability to generate CT images.
Source facts
Action
Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated January 2, 2018 to their customers. The letter identified the affected product problem and actions to be taken. For questions contact your local Philips representative or local Philips Healthcare office.
Initiated
2018-01-03
Root cause
Under Investigation by firm
Terminated
2019-07-02
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-0840-2018
Recall status
Terminated
CDRH record ID
161425
FDA record created
2018-03-02
Recall event number
79134