Skip to content

Public FDA record

Z-0840-2019: Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml G

Recall record Z-0840-2019 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0840-2019
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Feb 15, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

Source facts

Action
Randox the US distributor contacted customers by email on 12/4/18 to advise of problem and provided an update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments.The instrument testing order should be reviewed in line with the updated technical bulletin. A revised letter issued 12 Feb 2019 to advise "reagent would be observed as inconsistencies in Quality control recovery which may lead to a delay in running patient samples or erroneous elevated / depressed test results".
Initiated
2018-12-04
Root cause
Device Design
Terminated
2020-06-18
Distribution
WV, PR
Recall number
Z-0840-2019
Recall status
Terminated
CDRH record ID
169692
FDA record created
2019-02-15
Recall event number
81746

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗