Open original FDA record ↗Recall record
- FDA identifier
- Z-0841-05
- Company in FDA record
- Medrad Inc
- FDA classification date
- May 13, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Customer complaint that a continuum infusion pump was attracted to an MR scanner due to customer not completely engaging the pump into the bracket.
Source facts
- Action
- The recalling firm issued a Field Correction letter dated 3/9/05 to all their customers. The letter instructed the consignee on how to apply the new warning labels, to discard the old Operation Manual and replace it with the new one.
- Initiated
- 2005-03-22
- Root cause
- Other
- Terminated
- 2005-08-17
- Distribution
- The products were shipped nationwide to hospitals and outpatient MRI office. The product was shipped to government facilities in CA, DC, MD, NJ, SC, WA, and WV. The product was also shipped to Austriala, Canada, Mexico, United Arib, Austria, Germany, Denmark, France, Great Britain, Hong Kong, Iceland, Italy, Netherlands, Norway, Portugal, Sweden, and Singapore.
- Recall number
- Z-0841-05
- Recall status
- Terminated
- CDRH record ID
- 38165
- FDA classified
- 2005-05-13
- Recall event number
- 31565