Z-0841-05: Continuum MR Compatible Infusion Pump. Pump Model #3009135

Recall record Z-0841-05 for Medrad Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0841-05
Company in FDA record
Medrad Inc
FDA classification date
May 13, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Customer complaint that a continuum infusion pump was attracted to an MR scanner due to customer not completely engaging the pump into the bracket.

Source facts

Action
The recalling firm issued a Field Correction letter dated 3/9/05 to all their customers. The letter instructed the consignee on how to apply the new warning labels, to discard the old Operation Manual and replace it with the new one.
Initiated
2005-03-22
Root cause
Other
Terminated
2005-08-17
Distribution
The products were shipped nationwide to hospitals and outpatient MRI office. The product was shipped to government facilities in CA, DC, MD, NJ, SC, WA, and WV. The product was also shipped to Austriala, Canada, Mexico, United Arib, Austria, Germany, Denmark, France, Great Britain, Hong Kong, Iceland, Italy, Netherlands, Norway, Portugal, Sweden, and Singapore.
Recall number
Z-0841-05
Recall status
Terminated
CDRH record ID
38165
FDA classified
2005-05-13
Recall event number
31565

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗