Open original FDA record ↗Recall record
- FDA identifier
- Z-0841-2015
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Jan 13, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).
Source facts
- Action
- Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.
- Initiated
- 2014-09-15
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2015-05-20
- Distribution
- Worldwide Distribution
- Recall number
- Z-0841-2015
- Recall status
- Terminated
- CDRH record ID
- 131285
- FDA classified
- 2015-01-13
- Recall event number
- 69660