Z-0841-2018: IQon Spectral CT with 4.7.2 software version.- Model 728332, a Philips Computed Tomography X-Ray System

Recall record Z-0841-2018 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0841-2018
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA record creation date
Mar 2, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

23 Software related issues for the IQon Spectral CT with 4.7.2 software version. These issues include: " Software errors that may result in CT spectral images misrepresentation or incorrect treatment; " Software errors that may result in partial set of images or inability to generate CT images.

Source facts

Action
Philips Healthcare is notifying the affected users of these issues via a Field Safety Notice and will be releasing a software fix via Field Change Order to correct these issues. They also provide the following instructions.: Issue 1, Uric acid application for gout provides inconsistent results - When reviewing images using the Uric Acid Spectral Result type inconsistent results were provided on gout studies. The tendons would occasionally be displayed as containing Uric Acid even with no or normal uric acid blood levels. Issue 2, Multi-phase Cardiac CTA Scans should include a planned conventional result for every planned spectral result - When planning a Cardiac multi-phase study, it is possible to create a phase that consists of only a Spectral Result. Issue 3, 0-MAR information is missing in DICOM when saving as original or Secondary Capture - When an SBl that was created with 0-MAR enabled is used to create and save any Spectral result: " DICOM information about 0-MAR is missing but the 0-MAR label appears on a result saved as Original " DICOM information and the 0-MAR label are both missing on a result saved as Secondary Capture Issue 4, Attenuation plot presents incorrect HU values for MonoE - When comparing HU values of an ROl to the values in the attenuation plot graph when using the Show spectral plots for this ROI option, there is a difference of 10-20 HU values between the ROI HU value and the value shown on the graph Issue 5, Windows Operating System Time Zone Updates - Microsoft Windows Daylight Savings Time Zone updates have not been applied since December 2014. This may cause the system time to be incorrect at the start and end of daylight savings time. Issue 6, Sending Calcium Score findings to report using Mass Score protocol provides confusing information - The Heartbeat CS (Calcium Scoring) option provides confusing information in the Impressions section of the Calcium Scoring report. The report provides m
Initiated
2018-01-11
Root cause
Under Investigation by firm
Terminated
2019-05-20
Distribution
Domestic and Foreign
Recall number
Z-0841-2018
Recall status
Terminated
CDRH record ID
161445
FDA record created
2018-03-02
Recall event number
79148

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗