Z-0841-2020: KaVo Scan eXam SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

Recall record Z-0841-2020 for Palodex Group Oy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0841-2020
Company in FDA record
Palodex Group Oy
FDA record creation date
Jan 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.

Source facts

Action
On January 7, 2020 the firm distributed Urgent Medical Device Recall Communication letters by letter stating: Urgent Medical Device Correction Dear KaVo Kerr Customer, Palodex Group OY has identified an issue with Soredex DIGORA Optime, KaVo Scan eXam, Instrumentarium Express, and Gendex GXPS-500 Imaging Plate Systems (later called IPS). According to our records, we have delivered to you one or more devices belonging to the serial number range defined in the table below. The Power Supply Unit (PSU) delivered with the device may have a fault requiring correction. Product Soredex DIGORA Optime: Serial Number Range SJ1105219 to SJ1314421 KaVo Scan eXam: Serial Number range KJ1208178 to KJ1314403 Instrumentarium Dental Express: Serial Number Range IJ1208217 to IJ1313309 Gendex GXPS-500: Serial Number Range GJ1209488 to GJ1311722 Devices falling into these serial number ranges were delivered with PSUs having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX Please Note: If the PSU serial number does not begin with 1143, 1221, 1224, or 1229, then no action is required. In this letter we refer to DIGORA Optime, but instructions are applicable to all four makes listed in the table above. Issue Description Slow degradation of plastic insulation used in the PSU DC plug may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail. Risk of fire or injury has been assessed to be low. As a result, Palodex Group OY is being proactive with our customers to have these identified Imaging Plates Systems corrected. Please communicate this information to your local distributors and report the initiated Medical Device Correction to your local authorities if the local regulation requires distributors to do so. Palodex Group Oy as a manufacturer of the Imaging Plates Systems will take care of manufacturers reporting t
Initiated
2020-01-07
Root cause
Component design/selection
Terminated
2024-04-15
Distribution
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Recall number
Z-0841-2020
Recall status
Terminated
CDRH record ID
178572
FDA record created
2020-01-24
Recall event number
84544

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗