Z-0841-2022: Ysiosystem with VC10 software version, Model Number 10281013. Radiographic system

Recall record Z-0841-2022 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0841-2022
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Mar 30, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement. If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION dated 1/31/22 was sent to customers. What is the impact on system operation and what are the potential risks? If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons. What steps can the user take to avoid the potential risks associated with this issue? It is highly improbably that the user will experience the described issue. However, it is recommended to follow regular cautions and instructions listed in the system operator manual, particularly in the chapter Safety , section Information about unit movements . Users and all accompanying personnel are advised to remain outside the zone of potential device movement, e.g., potential movement area of longitudinal and transversal direction of the ceiling stand. How will the corrective action be implemented and how effective is the corrective action? Siemens Healthineers is distributing a Customer Safety Advisory Letter XP046/20/S to all affected customers. A field software update for Ysio units that will eliminate the described issue will be distributed via an Update Instruction XP045/20/S. Apart from the described correction, the updated software contains further stability, usability, and performance improvements. This resolution will be provided free of charge and is planned to be available in the first quarter of calendar year 2022. Your local Siemens Healthineers Service organization will contact you as soon as the software update is available for your system. Please feel free to contact our service organization for an earlier appointment at 1-800- 888-7436. Following the corrective action, the cause is eliminated and any recurrence of this potential fault is prevented Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisor
Initiated
2022-01-31
Root cause
Software design
Distribution
US Nationwide.
Recall number
Z-0841-2022
Recall status
Open, Classified
CDRH record ID
192270
FDA record created
2022-03-30
Recall event number
89703

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗